|
Plain Plug and Ring Gage Specifications
The
question often arises regarding tolerances on Plain Plug and Ring
Gages. The two main issues addressed in our calibration workshops
center around:
-
Tolerances:
Many people seem confused on what the tolerance should be or where to
find the actual required tolerances on Plain Plug and Ring Gages.
-
Laboratory Tolerance Requirements:
Complaints from people who say that The Calibration Laboratory
calibrates and reports their gages using tolerances that are too
tight or strict. Some of these companies do not hold their
tolerances so tight that they feel they need a tolerance held to
millionths of an inch.
Let's
address these issues one at a time.
-
Regarding
the tolerance requirements: Information for tolerances can be found
in ANSI B1.2 or in The Machinery's Handbook. It is listed as
American National Standard Tolerances for Plain Cylindrical Gages.
The following chart is from the Machinery's Handbook (disclaimer:
You should reference The Machinery's Handbook or the ANSI B1.2
standard. We do not recommend using this document as a final
decision.)
Size Range Tolerance
Class
Above to and including XX X Y Z ZZ
0.020 0.825 .00002 .00004 .00007 .00010 .00020
0.825 1.510
00003 .00006 .00009 .00012 .00024
1.510 2.510
.00004 .00008 .00012 .00016 .00032
2.510 4.510
.00005 .00010 .00015 .00020 .00040
4.510 6.510
000065 .00013 .00019 .00025 .00050
6.510 9.010
00008 .00016 .00024 .00032 .00064
9.010 12.010
00010 .00020 .00030 00040 .00080
2. The calibration Laboratory meets or exceeds the requirements of
one or more of these standards: ISO 10012, ISO/TS 17025,
ANSI/NCSL-Z-540, etc. The Laboratory is required to calibrate your
gages to a recognized standard. This requires them to calibrate your
plain cylindrical gages to the ANSI B1.2 standard.
CALIBRATION TIP
Indicator Calibration
Whether you are
calibrating a Test Indicator, Dial Indicator or Electronic Indicator,
one of your first checks should be a visual check of the contact point
to identify any worn or "flat" spots on the contact point. A worn or
"flat" contact point will cause the indicator to produce
non-linear readings. It is a simple check but one often over-looked.
Non-linear readings
can also be the result of improper contact point angles. Most test
indicators are designed to use parallel to the measured surface.
However, if you are using an Intertest Test Indicator, there is a 12°
Angle required between the contact point and the work piece. If you
are using a Best Test indicator there is a 10° angle required between
the contact point and the work pierce. I strongly suggest you read any
literature or special instructions that come with your indicator to
determine what angle is required. If the literature is not available
then check the catalogue or check with the manufacturer for proper
specifications .
This
calibration tip is written by Carlis Stuber, primary trainer for our Gage
Calibration and Hands-on Workshop. If you have any questions regarding this
article, please call 952-895-1150
Comment on this article

|